EC-3H
Diagnostic ECG Holter System
|
|
The
EC-3H device records three independent ECG channels with high
precision and detects pacemaker beats as well. The analysis
program performs automatic QRS detection, classification,
and rhythm analysis that can be interactively modified by
the user. The analyzing program performs time and frequency
based HRV, pacemaker, QT, AF and ST analysis of the 24 hours
recordings. The transparency and the interpretation of the
results are aided by the color coded graphs and tables. The
decision criteria may be changed, the classified beats and
event segments may be redefined. The quick and precise evaluation
of the recordings is helped by the efficient decision making
algorithms, the availability of a native language version
and the option for the generation of simple summarizing reports
make the system easy to use.
|
| |
3
independent + PM CH, ECG recording with high precision
|
| |
digital
signal storage in the solid-statememory for 24 hours
|
| |
small
size, light, compact built, convenient to wear
|
| |
option of patient event reporting in case of complaints
|
| |
economic
maintenance with rechargeable batteries
|
| |
simple user friendly operation surfaces
|
|
|
|
| |
QRS
classification and rhythm analysis
|
| |
time
and frequency based HRV, ST, QT, AF analysis with color
coded display
|
| |
Pacemaker
detection and analysis
|
| |
high
speed safe computer connection via IRDA/USB interface
|
| |
efficient
and user friendly analyzing and display programs
|
| |
PC
based evaluation in Win 98/ME, NT4.0, Win 2000/XP setting
|
|
|
|
Technical
features:
| |
ECG
channels: 3 independent + PM
|
| |
Dynamic
interval: +/- 9 mV
|
| |
DC
offset range: +/- 350 mV
|
| |
Input
impedance: min. 20 Mohm
|
| |
Frequency
response: 0.05-100 Hz
|
| |
PC
connection: high speed IRDA/USB (4Mb/sec)
|
| |
Average
data transmission time: 80 sec
|
| |
Power
supply: 2 x 1.2 V rechargeable batteries
|
|
|
|
|
|